AAS-7000 Series
The AAS-7000 series atomic absorption spectrophotometer is widely used in various fields like gold, petrochemical, geology, medical, environmental protection, scientific research, agriculture, disease control, food, materials science, commodity inspection and so on. At present, the instruments can analysis more than 70 kinds of elements. It play a very important role in trace elements’ analysis.
The AAS-7000 series atomic absorption spectrophotometer is widely used in various fields like gold, petrochemical, geology, medical, environmental protection, scientific research, agriculture, disease control, food, materials science, commodity inspection and so on. At present, the instruments can analysis more than 70 kinds of elements. It play a very important role in trace elements’ analysis.
Model | Configuration | Type |
AAS-7003 | Hose, flame, graphite furnace | Automatic |
Graphite furnace atomic transition temperature can reach 3000 ℃, in line with national standards for nickel, molybdenum, vanadium, cobalt and other elements of high-temperature graphite furnace atomic temperature requirements.
The advanced optical systems ensure the instrument's high optical power, when detect with graphite furnace, the negative pressure lower about 100 ~ 200V than other domestic products under the same conditions (each element can reach this level), high signal to noise ratio ensure a stable data.
The specially designed graphite furnace increases the maximum injection volume to70 μL, which is conducive to multiple injections and low concentration samples.
Equipped with continuous light (D2), self-absorption background correction system, which can correct up to 1.5A of background interference.
The heating rate is greatly increased due to the use of optical heating rate, fast atoms, improve analytical sensitivity and accuracy.
The unique graphite furnace design ensures atomic effects of the isothermal time, for obtain accurate data to provide a guarantee.
Uses high-tech aerospace, lost wax casting process titanium spray chamber and burner head. Has excellent resistance to corrosion, oxidation, aging, and high temperature, long service life;
The design of sample inlet reduces the sample error, which make sample injection easier, and ensure that even a manual injection can achieve good accuracy.
Graphite cone can be easily replaced after wearing down, to ensure the stability of the electrode conductivity, greatly improving the efficiency of instrumental analysis. (Patent No.: ZL2007 2 0104071)
Based on Windows XP platform, multi-language support, such as English, Korean, etc., can quickly and easily draw analysis. Powerful analysis capabilities and fault-tolerant samples, flexibility to modify the curve fitting equation
Flexible report printing, users can print according to the different choice.
To achieve the management of history and related follow-up treatment, including query, editing and other functions. Can download a variety of file formats and data files.
Automatically determine whether the result of the analysis is appropriate, if out of range, the system automatically re-measured in accordance with set requirements. About the system's QA / QC features include: standard deviation (SD) test, the relative standard deviation (RSD) detection, correlation coefficient testing, QC testing, baseline drift (sensitivity correction) test, the sample detection limit (automatic on-line dilution).
Automatic element lamps selecting, automatic wavelength scanning, automatic slit switching, lamp current automatically set, etc.
Real-time dynamic monitor equipment working status of each component. Flame: to monitor the combustion head type (normal or high-temperature combustion head), to monitor water seal level, gas, the auxiliary gas switch, pressure, auxiliary gas current, acetylene gas leakage alarm. Graphite furnace: over-current protection, monitoring water temperature, water flow, gas pressure monitoring and protection.
automatically record the time of element light.
Wavelength range | 190 ~ 900nm |
Monochromator | C-T grating monochromator |
Wavelength repeatability | ≤0.1nm (AAS-7002, AAS-7003) |
blaze wavelength | 250nm |
Resolution | better than 0.3nm |
Spectral Bandwidth | 0.1, 0.2, 0.4, 1.0, 2.0 nm five-speed auto switching (AAS-7002, AAS-7003) |
Wavelength accuracy | ± 0.2nm |
Grating grooves | 1800 / mm |
Baseline stability | ≤ ± 0.004A/30min (dynamic), |
Cu characteristic concentration | ≤ 0.02 μg / ml |
Precision RSD | ≤ 1% |
Position adjustment | Best regulation for height and angle, replace Flame / hydride in 1min. |
Detection limit | ≤ 0.005 μg / ml |
Combustion head | 50* 100mm interchangeable all-titanium burner head, all-titanium spray chamber |
Cd features volume | 0.5×10-12g |
Temperature range | room temperature~3000 ℃ |
Temperature program | Temperature program up to 20 bands, ladder, ramp, Maintain the three kinds of heating methods. |
Heating modes | maximum power heating and light-controlled rapid heating |
Detection limit | 0.5×10-12g |
Precision RSD | ≤ 2% |
Ramp rate | ≥ 2000 ℃ / sec. |
Flame and graphite furnace background correction can be all achieved.
Correction methods: deuterium lamp, self-absorption background correction (optional)
Correction capability: 1.5A background correction can be corrected, correction capability> 50-fold.
Measuring method: flame absorption, flame emission method, graphite furnace, hydride method.
Analysis: linear equations, nonlinear equations, the standard addition method. Print output: the calibration curve, the signal spectrum, instrument conditions, analysis parameters and results can be automatically stored and printed.
Size: 880 (L) × 540 (width) × 450 (high) mm, 125kg.
Power supply: 220V, 50Hz, single-phase
Host: 200W, graphite furnace 4KW
Main Features
Hydride generator
Atomic absorption flame auto sampler
Graphite furnace auto sampler