CLG-Max 3500FL
CLG-Max 3500 Multifunction fluorescence microplate reader suitable for life science research, with light absorption, upper fluorescence detection, bottom fluorescence detection, chemiluminescence and other detection mode, it is the most powerful research fluorescence plate reader
Multifunction Fluorescence Microplate Reader
Model: CLG-Max 3500FL
CLG-Max 3500 Multifunction fluorescence microplate reader suitable for life science research, with light absorption, upper fluorescence detection, bottom fluorescence detection, chemiluminescence and other detection mode, it is the most powerful research fluorescence plate reader .It s capable of detecting any standard 96-1536-holesl microplates. Built-in grating excitation monochromator, continuous wavelength range of 190-1000nm, stepping wavelength 1nm.It is suitable for light absorption, fluorescence detection, chemiluminescence detection and a variety of ways.
Main Application
DNA / RNA / protein quantization and purity testing
Endpoint ELISAs / EIA Cell proliferation / cytotoxicity
IC50 / LD50 protein quantitative analysis (BARADFORD, BCA,Lowry)
Immunoassay analysis Drug dissolved analysis
Enzyme kinetic analysis ADME drug analysis
LAL endotoxin analysis Cell detection (MTT, etc.)
Identification of bacterial Calcium Assay
chemotaxis penetrate the cell membrane experiment
green fluorescent protein experiments Molecule probe experiment
Membrane potential detection Oxidation reaction
Livpotential detection oxidation detect Live / Dead cell viability / Cytotoxicity Assay
NanoOrange protein experiments Caspase-3 Apoptosis Detection PicoGreen DNA detection Luciferase analysis
CatchPoint cAMP Detection Dual Luciferase analysis
Tyrosine Kinase Assay dsDNA ELISAs
Aequorin Ca? + Analysis Based chemiluminescence analysis
Features
1, Grating emitting monochromator ,continuous wavelength 190-1000nm, no need for additional filters;
2, With a strong light absorption detection function, can be used as ultra trace ultraviolet-visible spectrophotometer;
3, With no fiber optics, optical path variable, small cross-impact, with high sensitivity;
4, Fluorescence detection can be at the top , fluorescence detection can also be carried out at the bottom, with a strong fluorescence analysis ability.;
5, Fully closed cabinet light, using a photon counting photomultiplier tube for chemiluminescence analysis;
6, Wth multi-point detection average technique, may well overcome the problems of not accurate detection which caused by irregularity of samples
7, The use of advanced virtual thermostat cover technology that can precisely control the temperature of the sample compartment;
8, Has a powerful data analysis and instrument control software, highly automated;
9, Provides endpoint detection, dynamic, multi-wavelength, full wavelength scans detected in four read modes;
10, Can be added with auto sampler, accurate reagent is added to meet the rapid chemiluminescent reaction;
Parameters
1, The excitation wavelength range: 190nm-1000nm;
2, The range of emission wavelength: 220nm-1000nm;
3, The light absorption range: 190nm-1000nm;
4 Wavelength selection: grating monochromator, adjustable 1.0nm step
5, The excitation wavelength band: 2-20nm adjustable;
6, Wavelength accuracy: <± 0.2nm;
7, Measurement range: 0-4.600OD;
8,Linear range: (405nm) 0-3.50OD;
9, Microplate readings Time: Endpoint: 10 seconds Kinetics: 0.1 seconds Minimum interval
10 ,Microplates: 6-1536 hole;
11, The vibration plate functions: linear, circular, cross shock;
12, Temperature range: room temperature ~ 65 ℃;
13, Detector: Photomultiplier tubes (transmission, reflection);
14,Means of communication: USB;
15,Detection method: light absorption, fluorescence detection at the top, bottom fluorescence detection, chemiluminescence;
Main components
1, Host
2, CLG-pro3500FL analysis software
Options
1,Sample: for rapid kinetics reaction; can add with 1-4 channels;
2, SP-100 type vibration plate - incubation - Dosing all-in-one machine;