PT-D92-3536-1
无标题文档
Summary
This instrument is designed and made as per GB/T3536-2008 Petroleum Products - Determination of Flash and Fire Points - Cleveland open cup method and ASTM D 92 Standard Test Method for Flash and Fire Points by Cleveland Open Cup Tester. It is applicable to all petroleum products with flash points above 79°C and below 400°C except fuel oils.
Main technical features
Main technical specifications
1. Power supply: AC (220±10%)V, 50Hz.
2. Heating device: Electric furnace heating, no naked fire, explosion prevented.The power is adjustable from 0W to 600W. The max heating temperature can reach 400°C.
3. Temperature control: Single chip microcomputer.The heating rate can meet requirements of standards GB/T 3536-2008 and ASTM D92.
4. Temperature display: LCD shows temperature parameters.
The display range is 0°C ~ 400°C, display accuracy is 0.1°C.
5. Flash point detecting device: It applies the test flame automatically.
6. Temperature sensor: RTD, PT100
7. Igniting device: (1) Ignition source: coal gas (or civil gas)
(2) Flame diameter is 3.2mm ~ 4.8mm
8. Ambient temperature: (-10~50)°C
9. Relative humidity: ≤85%
10.Maximum power consumption: 650W
11.Dimension: 340mm×320mm×450mm (Temperature sensor is included)
Standard Accessories
No. | Item | Unit | Qty | Remarks |
1 | Oil sample cup | Piece | 1 |
|
2 | Igniting gun | Piece | 1 |
|
3 | 5A fuse (Ф5×20) | Piece | 2 |
|
4 | Latex pipe Ф3 0.25m | Piece | 1 |
|
5 | Latex pipe Φ6 | Meter | 1.5 | This pipe is not suitable to be used as gas pipe. |
6 | Flameout cover | Piece | 1 |
|