CIT20-10S Koenen test instrument used to determine the sensitivity of the solid and liquid substances under high confinement effect of heat. The instrument complies with United Nations "on the Transport of Dangerous Goods Recommendations on Tests and Criteria Manual". Section 11.5.1 Test 1 (b) Koenen test.

It equipped with a standard steel plate at one end of instrument, after loading sample into pipe, automatic electronic ignition and heating the steel pipe, it using high-precision timing system, heating automatic stop. Automatic pop-up report after the end of the test, user can observe results and select from the report, and can determine the thermal sensitivity of material as per steel deformation or rupture extent.

Product Features
1) Embedded processor, Windows CE operating system;
2) 8-inch LCD touch screen which display the test status in real time;
3) Integration "standard model" and "Calibration Mode"  to meet user test requirements;
4) Auto-ignition heating system functions
5) 80N constant pressure automatic specimen preparation, precise sample preparation and improve the test results;
6) The device can be realized the heating rate control through the flow meter;
7) Automatic heating & timing system to achieve the integration of test operation;
8) Front door openable to convenient test results viewing which is fast and clean;
9) Equipped with video monitor system to record the testing process, the final result was observed after the test results through playback, remote monitoring and judgment;
10) Remote control ignition;
11) Testing process fully automated, no need of human intervention, support parameter setting, data storage, viewing reports.

Product parameters
1) Operating environment: 5 ℃ ~ 30 ℃, <85% RH
2) Temperature measuring range: 0 ~ 500 ℃
3) Temperature Measurement Accuracy:
     Below 300 ℃ of ± 2.5 ℃
     Above 300 ℃ of ± (0.75% of reading)
4) sample preparation Pressure: 80N
5) sample preparation ballast weights Accuracy class: F2
6) heating rate: 3.3 ± 0.3k / s
7) Steel weight: 26.5g ± 1.5g
8) pipe wall thickness: 0.5mm ± 0.05mm
9) orifice aperture: 1.0mm, 1.5mm, 2.0mm, 2.5mm, 3.0mm, 5.0mm, 8.0mm, 12.0mm, 20.0mm
10) pipe size Φ (outer diameter) × H: 25mm × 75mm