Hazardous Chemical Detection

Aerosol spray enclosed space tester

CIT10-10S

 CIT10-10S aerosol spray enclosed space test instrument is used to assess flammability of the aerosol in a closed or restricted space

More details

More info

※product description:
   CIT10-10S aerosol spray enclosed space test instrument is used to assess flammability of the aerosol in a closed or restricted space. It is in line with the United Nations "on the Recommendations on the Transport of Dangerous Goods • Manual of Tests and Criteria" 31.5 enclosed space ignition test and GB / T 21631-2008 "dangerous sprays enclosed space ignition test methods."
     Instrument consists of two parts, the display control section, the detection system composed of a cylindrical container body detection system, built-in light sensors and high-precision thermocouple sensor automatically determine whether knocking occurred. The operator can monitor the entire testing process through the display control section LCD screen.

※ Product parameters:
1) Working environment: 5 ℃ ~ 30 ℃, <85% RH
2) Ambient temperature measuring range: -10 ℃ ~ + 60 ℃
3) the ambient temperature measurement accuracy: ± 0.5 ℃
4) Humidity measuring range: 5 ~ 95% RH
5) Humidity measurement accuracy: ± 3.0% RH
6) combustion detection response time: ≤6ms
7) pressure measurement (differential pressure) range: -5 ~ 5kPa
8) Pressure measurement accuracy: ± 0.1% FS
9) Timing accuracy: ± 0.1s
10) ignition flame: standard candle flame, adjust the pedestal height 20mm / a;
11) cylindrical vessel size: Ф600mm × 720mm
12) aerosol cans Location: aerosol tank automatically flips
13) compatible spray cans Dimensions: Height 120 ~ 400mm, diameter of 30 ~ 70mm
14) The distance from the cylindrical vessel to the spray nozzle: 30-35mm adjustable
15) effusion plate: spray can accumulate discharged residual liquid, it can be withdrawn and processed at any time .

※Product Features:
1) embedded processor, Windows CE operating system;
2) 8-inch LCD touch screen, display the test status in real time;
3) integration of the "standard model" and "Custom Mode" in two modes;
4) Spray automatic sprays and spray cans clamping device can automatically make the tank vertical and inverted;
5) pressure sensors, nanosecond light sensors double detection module, automatically detects whether a flame is burning or explosion in a closed space;
6) added to the ventilation system, the smoke exhaust from the barrel, to ensure test accuracy and safety of laboratory personnel;
7) testing process fully automated, without human intervention, support parameter setting, data storage, viewing reports, etc;
8) equipped with temperature and humidity module, real-time monitoring the temperature, humidity of test environment .