Hazardous Chemical Detection

Self-heating substances tester

CIT01-10S

CIT01-10S self-heating substances tester is used to determine whether the substance oxidize and self-heat exposed in air under a certain temperature

More details

More info

※product description:
      CIT01-10S self-heating substances tester is used to determine whether the substance oxidize and self-heat exposed in air under a certain temperature, in line with the United Nations "on the Recommendations on the Transport of Dangerous Goods, Tests and Criteria Manual" section 33.3.1.6 Test N. 4 test method for self-heating substances.
     Instruments is the desktop box structure, consist by the display control section and the heating system two parts, with a side length of 25 mm, 100 mm cubic steel mesh container for placing detection sample. Heating system designed removable interior layers and hot air circulation system, front-mounted explosion-proof glass box observation window for easy observation of the sample state test. The instrument can continuously detect the internal temperature of the sample within 24 hours, generate data reports, and automatically determine the test results in accordance with the appropriate standards.
※ Product parameters:
1) Working environment: 5 ℃ ~ 30 ℃, <85% RH
2) Temperature range: room temperature ~ 200 ℃
3) Temperature accuracy: ± 1 ℃
4) experimental area average temperature: Below 140 ℃ of ± 1.5 ℃
5) Temperature measuring range: 0 ~ 500 ℃
6) Temperature measurement resolution: 0.01 ℃
7) Temperature measurement accuracy: below 300 ℃ is ± 2.5 ℃
8) standard time range: 0 ~ 24h
9) Custom time range: 0 ~ 48h
10) cubic steel mesh containers: small container: the length of 25mm, aperture 0.05mm;
      Small container housing: 30mm × 30mm × 30mm;
      Large containers: the length 100mm, aperture 0.05mm;
      Large container cover: 105mm × 105mm × 105mm;
      Cage size: 150mm × 150mm × 250mm;
11) Liner Size: 350mm × 350mm × 400mm

※Product Features:
1) embedded processor, Windows CE operating system;
2) 8-inch LCD touch screen, display the test status in real time;
3) Built-in hot air circulation heating system, heating uniformity and ensure stability;
4) digital PID algorithm to precisely control the temperature of heating, temperature control accuracy up to ± 1 ℃;
5) Test report contains package type of judgment;
6) automatic fault detection instruments and sound and light alarm, ensure that the test is more secure;
7) Test can be diversified; the user can configure the test parameters based on characteristics of the sample, arbitrary adjustment test sequence;
8) cage with special specifications can be customized;
9) real-time display running time, real-time temperature, automatic drawing test curve, you can view the raw data;
10) the front of the device provided with explosion-proof glass windows, the back pressure relief door design, internal thrust greater than 2.5 kgf, relief doors open, ensure that the test is more secure;
11) testing process fully automated, without human intervention, support parameter setting, data storage, viewing reports.

30 other products you may be interested: