BSC Series
Biological safety cabinet is indispensable equipment for modern microorganism lab, especially for those occasions that the operators need to take protective measure, such as medicine, pharmacy, microbiology, biomedicine, gene recombination, animal experiment and biological products. It can provide a sterile clean and safe environment.
Application Range:
Biological safety cabinet is indispensable equipment for modern microorganism lab, especially for those occasions that the operators need to take protective measure, such as medicine, pharmacy, microbiology, biomedicine, gene recombination, animal experiment and biological products. It can provide a sterile clean and safe environment.
Structural Characteristics:
Technical Parameters:
Model | BSC-1000IIA2 | BSC-1300IIA2 | BSC-1600IIA2 | ||
Exhaust capacity | 30% | 30% | 30% | ||
Clean level | ISO 5(Class 100)/ISO 4(Class 10) | ||||
HEPA Filter Efficiency | @0.3~0.12µm≥99.999% ~ 99.995% | ||||
Downflow Air Velocity | 0.25m/s~0.5m/s | ||||
Inflow Air Velocity | ≥0.5m/s | ||||
Noise | ≤60dB(A) | ||||
Vibration Half Peak | ≤5µm | ||||
Electric Power | AC, Single-phase 220V/50Hz(AC,220V,1¢,50Hz) | ||||
Laminar Flow Protection | Operator Protection(Repeat 3 times) | a.Total colony in impaction sampler≤10CFU/time | |||
Sample Protection(Repeat 3 times) | Total colony in culture dish≤5CFU/time | ||||
Protection Against Cross-Contamination(Repeat 3 times) | Total colony in culture dish≤2CFU/time | ||||
Maximum Power | ≤1500KVA(with spare socket) | ||||
Weight(kg) | 150 | 200 | 250 | ||
Operation Room Dimension(mm) | 1000×570×635 | 1300×570×635 | 1600×570×635 | ||
Overall Dimension(mm) | 1200×855×2350 | 1500×855×2350 | 1800×795×2350 | ||
Fluorescent Lighting | 30W×(2) | 40W×(2) | 40W×(2) | ||
UV Lighting | 20W×(1) | 30W×(1) | 40W×(1) | ||
Luminance | ≥650Lx | ||||
Exhaust Pipe’s Diameter | Φ200 | Φ250 | |||
Exhaust Direction | Top ejection |