SP-F97/SP-F97pro
SP-F97/SP-97pro Fluorescence Spectrophotometer, Spectrofluorometer
SP-F97/F97pro fluorescent spectrophotometer is a new generation of high performance molecular luminescence analysis instrument. The product structure is exquisite, has the characteristics of high detection sensitivity, fast scanning speed, wide spectrum measuring range, high dynamic range, fast 3D scanning, and so on. Easily meet the requirements in the field of material research, pharmaceutical analysis, biochemical and clinical testing, water quality analysis and control, food safety testing (dairy products, aquatic products, such as vitamin C, selenium, aflatoxin), and other areas.
Features:
High detection sensitivity: based on high efficiency optical design and weak signal detection technology, the water Raman peak signal to noise ratio can be greater than 200 (P - P) to the leading domestic and international advanced level.
High scanning speed: the high speed digital signal processing unit provides the world's fastest scanning speed at 48000nm/min. Only 1 second to get classic fluorescence spectra, 1 minute to get high quality of three-dimensional fluorescence spectra.
Wide Spectral measurement range: using a double monochromator design, excitation and emission continuous spectral range covering 200 nm to 900 nm, meet the needs of most of the fluorescence analysis of the test.
Excitation light path monitoring system: instrument is equipped with excitation light dual beam ratio monitoring system to ensure the fluorescence signal high and stable.
High quality assurance: using Hamamatsu’s high quality longevity life Xenon light source and photoelectric multiplier tube detectors and instruments to provide sufficient light intensity signal and the detection sensitivity.
Built-in optical gate: built-in programming control optical gate device, specifically for prone to reaction to light or fluorescence signal unstable sample design, through a unified excitation detection to improve the accuracy of fluorescence analysis.
Specification
Model | SP-F97 | SP-F97pro | SP-97xp |
Detection sensitivity (water Raman peak signal noise ratio) | S/N ≥150 (P-P) | S/N ≥200 (P-P) | |
Excitation wavelength Bandwidth | 10nm | 2nm/5nm/10nm/20nm | |
Emission wavelength Bandwidth | 10nm | 2nm/5nm/10nm/20nm | |
Wavelength accuracy | ± 1.0nm | ± 1.0nm | ±0.4nm |
Wavelength repeatability | ≤0.5nm | ≤0.5nm | ≤0.2nm |
Light source | 150W automatic deodorizing xenon lamp (Hamamatsu, Japan) | ||
Optical system | Double-grating monochromator automatic wavelength scanning with optical excitation double beam ratio monitoring | ||
Fluorescence Receiver | Original photomultiplier (Hamamatsu, Japan) | ||
Excitation wavelength range | 200nm~900nm | ||
Emission wavelength range | 200nm~900nm | ||
Wavelength scanning speed | 10 positions, fastest 48000nm/min, fine scanning 15nm/min | ||
Measurement line | ≥0.995 | ||
Gain adjustment range | 1~17 position optional | ||
Integral Response time | 6 position options: 4s ~ 20ms | ||
Data output | Microsoft (R) Excel format, bmp image format | ||
Data transmission | USB2.0 Interface | ||
Light value range | 0.00-10000 | ||
Peak intensity Repeatability | ≤1.5% | ||
Power source | 220V / 50Hz, 110V / 60Hz | ||
Standard energy consumption | 200W | ||
Dimensions (mm) | 380×445×310 | ||
Weight | GW 14kg, NW 12kg |