TOC Analyzer

TOC-10 Portable TOC(total organic carbon) Analyzer (Offline Test&cleaning validation)

TOC-10 TOC

TOC-10 TOC tester is designed for the online and offline test of pharmaceutical grade water, WFI(water for inj.), ultra-pure water, and deionized water and etc. Based on the conductivity difference technology, it is being of high-precision capacity and short response-time, also meets the regulation standards issued by government.

More details

More info

Meet the requirements of TOC testing regulated by the Chinese Pharmcopoeia (2010 version) for pharmaceutical grade water

 

TOC-10 TOC tester is designed for the online and offline test of pharmaceutical grade water, WFI(water for inj.), ultra-pure water, and deionized water and etc. Based on the conductivity difference technology, it is being of high-precision capacity and short response-time, also meets the regulation standards issued by government.

 

Applications

Pharmaceutical water (purified water, water for injection) on-line monitoring and laboratory testing and cleaning validation;

Environmental testing, electronic industry, food industry and so on.

 

Working Principle

 

This tester adapts the UV-oxidation method,to oxide the sample organic contaminants into carbondioxide gasthoroughly which dissolve into liquid phase instantly,this will result the conductivity difference signal between test tube and control tube,and the final TOCvalue can be converted out by the below formula:TOC=TC(total carbon in test tube)- TIC(total inorganic carbon incontrol tube).

Features of product:

  1.       Lowcostsofoperating,thewholeprocessdoesnotrequiretheuseofanysupplies,suchasworkinggas,reagents,andotherparts.
  2.       Friendlyinterface,easyandconvenientoperation
  3.       Twomode,optionalonlineorofflinemodewiththechoiceofautomaticinjector.
  4.       Large capacity memory disk.
  5.       Portabledesign,suitableforuseinvarietyoccasions.
  6.       Automaticalarm,underthecircumstanceofsampleTOCconcentrationexceedinglimit.
  7.       Electronicsignature,aswellasusergrademanagement.
  8.       Good reproducibity of testing, supported by the high-precision sampling chip.
  9.       GMP & FDA certificated,meetswellwiththeregulationofChinesePharmacopoeia(2010version)issuedbygovernment.

 

Schematic Diagram of Direct Conductivity Method

1.Delay line

2. UVlamp and quartz tube

3.Conductivity sensor

4.Conductivity sensor

5.Pump

Specifications:


Model

TOC-10

Power requirements / power

100-240VAC, 50HZ, 120W

TOC detection range

1μg / L ~ 1500 μg / L

Conductivity detection range

0.055μs / cm-8.000μs / cm

Detection limit

0.001mg / L

Accuracy

≤3% test range

Repeatability error

≤ 3%

Zero drift

≤± 5%

Sample temperature

1- 99 ℃

Ambient temperature

5-65 ℃

Power / Frequency

100-240VAC / 50HZ; 120W

Dimensions

410 * 240 * 310mm

equipment weight

8.5kg

Display

Color touch screen

Resolution

0.001mg / L

Measuring range

0.001mg / L-1.5mg / L (1500ppb)

measurement accuracy

± 3%

Response time

Within 5 minutes

Scope of application

Offline laboratory, cleaning verification

Audit trail

≥5 years storage

authority management

Username & password login, level 4 permissions, meeting FDA 21 CFR PART 11

Print function

External mini printer

history record

≥5 years storage

data backup

Support U disk export data

Display

Color touch screen

Overall dimensions (mm)

400 × 240 × 270 (length × width × height)

PC software

able to provide

Function compare


Model

TOC-10

TOC-20

TOC-30

Temperature measurement

Online analysis

Offline analysis

Zero calibration

Adaptive verification

Pipe flushing

Pump tube replacement

Automatic decompression

Control output

Telecommunication

Wireless alarm

Audittrail

 

Auto sampler

 

● Standard        ◎Optional